Flexim Fluxus
ADM Ultrasonic Flowmeter
|
|
Measuring Principle
|
Ultrasonic Time Difference Correlation
Principle & Noise Trek
|
|
Measured Values
|
Volume Flow , Mass Flow , Heat Flow
, Flow Velocity , Sound Speed in Fluid
|
|
Totalizeres
|
Volume , Mass , Heat
|
|
Flow Velocity
Range
|
0.01 … 25 m/s
|
|
Resolution
|
0.025 cm/s
|
|
Repeatablity
|
0.15% of Measured Value
0.015 m/s
|
|
Accuracy
|
Volume Flow
|
±1 … 3% of Meas.Value (depending
on application) 0.015m/s
±0.5% of Meas. Value with Process Calibration 0.015 m/s
|
|
Path Velocity
|
±0.5% of Meas. Value
0.015m/s
|
|
Gaseous &
Solid Content
|
‹ 10% of Volume
|
|
Ambient Temperature
|
-10 … +60 ºC
|
|
Clamp-On
Flow Transducers(Type Q3N,Q3E)
|
|
Rated Pipe Diameter
Range
|
(10) 25 … 400(1000)mm
|
|
Temperature
Range
|
Q3N -30 … 130ºC
Q3E -30 … 200ºC
For Short Periods Up to 300ºC
|
|
Wall Thickness
Measurement
|
|
Measuring Range
|
1mm … 200mm
|
|
Resolution
|
0.01mm
|
|
Linearity
|
0.1mm
|
|
Temperature
Range
|
Standard Version
-20 … +60ºC
|